Vacuubrand DVR 2 User Manual Page 10

  • Download
  • Add to my manuals
  • Print
  • Page
    / 17
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 9
page 10 of 17
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. manual-no.: 999001 / 04/04/2012
Press ”ON/OFF” key to conrm the measuring cycle and to
terminate the mode.
When switching off the device, the remaining battery capac-
ity is displayed qualitatively by the number of arrows in the
display.
If the battery symbol appears in the display during meas-
urement, the battery should be replaced.
The correct operation of the device is no longer ensured.
Display of the battery‘s discharge status
Troubleshooting
Fault Possible cause Remedy
No display or display
disappears.
Automatic sleep mode?
Switch on, increase opera-
ting time if necessary.
Battery at?
Replace battery.
Other cause?
Contact local distributor.
Incorrect pressure read-
ing.
Pressure transducer not cor-
rectly adjusted?
Readjust pressure trans-
ducer.
Moisture in the pressure
transducer?
Dry or evacuate the pressure
transducer and readjust if
necessary.
Battery at, battery symbol is
displayed?
Replace battery.
Display is ickering and/
or battery symbol ap-
pears.
Battery at?
Replace battery.
Adjustment mode cannot
be activated.
At the applied pressure, read-
justment is not possible.
Admit air to atmospheric
pressure or evacuate to a
pressure below 20 mbar.
All segments of the LCD
are displayed or no dis-
play although battery has
been replaced.
Pressure transducer or meas-
uring electronics defective?
Contact local distributor.
Page view 9
1 2 ... 5 6 7 8 9 10 11 12 13 14 15 16 17

Comments to this Manuals

No comments