Vacuubrand RZ 9 User Manual Page 10

  • Download
  • Add to my manuals
  • Print
  • Page
    / 49
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 9
page 10 of 49
Make sure ventilation is adequate to maintain recommend-
ed operating temperature. Keep a minimum distance of
2 in (5 cm) between the cooling fan and surrounding items
(e.g., housing, walls, etc.), or else install an external auto-
matic ventilation system. Check fan regularly for dust/dirt.
Clean fan guard grill if necessary to avoid a reduction of
ventilation.
Use only hoses at the inlet and outlet of the pump with
an inner diameter at least as large as the diameter of the
pump’s connections (to avoid overpressure at the outlet,
and reduction of pumping speed at the inlet).
Allow the equipment to equilibrate to ambient temperature
if you bring it from cold environment into a room prior to
operation. Notice if there is water condensation on cold
surfaces.
Comply with all applicable and relevant safety require-
ments (regulations and guidelines). Implement the re-
quired actions and adopt suitable safety measures.
Ambient conditions
Do not reach for this product if it has fallen into liquid.
There is a risk of deadly electrical shock. Unplug the
system immediately.
+ Do not use this product in an area where it can fall or
be pulled into water or other liquids.
Adopt suitable measures in case of differences from
recommended conditions, e.g., using the equipment
outdoors, installation in higher altitudes, conductive
pollution or external condensation on the pump.
Do not operate this product near ames.
NOTICE