Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. Manual-no.: 999188 / 19/05/2009
Troubleshooting
Fault Possible cause Remedy
No display.
Mains not plugged in?
Plug in mains plug of pump-
ing unit.
Device switched off?
Switch on device.
VACUU • BUS cable to inter-
nal power supply not plugged
in at controller?
Plug in VACUU • BUS cable
at controller CVC 3000.
Other causes (device defec-
tive)?
Contact local distributor.
Display disappears.
Too much load (e.g. valves)
connected?
Check current draw of the
connected devices.
Short circuit at connected
valves?
Replace valves.
Short circuit at the RS 232
plug?
Check plug and cable.
Other causes (device defec-
tive)?
Contact local distributor.
Pressure reading incor-
rect.
Device not adjusted?
Readjust CVC 3000.
Humidity in the measurement
chamber?
Let the pressure transducer
dry, e.g. by pumping. Read-
just if necessary. Determine
and eliminate the cause for
humidity.
Pressure transducer contami-
nated?
See „Cleaning the pressure
transducer“.
Digital pressure read-
ing is ashing, ”0.0” is
displayed.
Pressure transducer not
correctly adjusted under
vacuum?
Adjust CVC 3000 correctly.
No digital pressure read-
ing.
Pressure transducer defec-
tive?
Contact local distributor.
Digital pressure reading
is ashing, one blip*.
Overpressure at the pressure
transducer, pressure > 1060
mbar?
Release pressure immedi-
ately (risk of bursting).
Warning triangle and
black valve symbol are
ashing, two blips*.
External venting valve re-
moved or defective?
Connect valve or replace by
a new one or recongure
without valve.
Warning triangle and
valve symbol are ash-
ing, three blips*.
NT VARIO / VARIO-B pump
and in-line valve connected?
Disconnect one component,
switch controller off/on.
Warning triangle and
coolant valve symbol are
ashing, four blips*.
Coolant valve removed or
defective?
Check connection cable of
the valve or use new valve or
recongure without valve.