Vacuubrand DCP 3000 User Manual Page 14

  • Download
  • Add to my manuals
  • Print
  • Page
    / 64
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 13
page 14 of 64
+ The equipment is designed for operation at ambient
temperatures and gas temperatures at the pressure
transducer between +50°F and +104°F (+10°C and
+40°C), or for short periods up to +176°F (+80°C) at
the pressure transducer (gas temperature). Periodical-
ly check maximum temperatures if installing the equip-
ment in a cabinet or a housing. Make sure ventilation
is adequate to maintain recommended operating tem-
perature. Install an external automatic ventilation sys-
tem if necessary. If processing hot gases, make sure
that the maximum permitted gas temperature at the
pressure transducer is not exceeded (see “Technical
data”, pg. 21).
Ensure that the wetted parts are chemically resistant
to the processed substances prior to operation.
Use the equipment only as intended, that is, for mea-
surement of vacuum in vessels designed for that purpose.
Any other use will automatically invalidate all warranty and
liability claims.
Remain aware of safety and risks.
Setting up and installing the equipment
+ Do not permit any uncontrolled pressurizing (e.g.,
make sure that pipelines cannot become blocked) to
avoid a risk of bursting!
+ Keep the VACUU•BUS cables away from heated sur-
faces.
Comply with maximum permissible pressures at the
pressure transducer. See section “Technical data”, pg.
21.
Connect hoses gas tight at the pressure transducer.
Secure hoses at hose nozzles (e.g., with hose clamp)
to prevent their accidental slipping. Ensure the stability
of hose connections.
NOTICE