The vacuum gauge DCP 3000 is equipped with a serial interface (RS
232C, nine-pin Sub-D-plug).
+ Plug-in or remove the cable (cable RS 232C) from the interface only if
the equipment is switched off.
+ The interface is not electrically isolated from the measuring circuit.
+ For optimal electromagnetic compatibility assemble an interface lter
(cat. no.: 638235).
+ The serial interface is prepared for the connection of a commercial
Bluetooth adapter for wireless communication (current supply for Blue-
tooth adapter on pin 9, see ”Pin assignment”).
+ Settings via interface are stored.
Interface parameters
Setting of the interface
Set the interface parameters directly at the DCP 3000. The factory set
values are underlined.
Edit and conrm the interface parameters in the ”Conguration” menu in
”RS-232” submenu using the selection knob.
Baud: 2400, 4800, 9600 or 19200
Parity: 8-N-1, 7-O-1 or 7-E-1
Handshake: Off, Xon-Xoff or RTS-CTS
Send: Off or 1......3600 s
Data: Send
A maximum of twenty commands per second is possible.
The commands have to be written in capital letters.
Command and parameter have to be separated by a blank.
The string is terminated with <CR> or <LF> or <CR><LF>.
The response of the controller is always terminated with <CR><LF>.
Numerical values and parameters can be written without leading zeros.
The response of the controller always includes leading zeros.