page 39 of 64
determined and indicated between two VSK 3000 pressure transducers
connected to the DCP 3000.
+ This requires one of the two pressure transducers to be dened as
reference (e.g., VACUU•BUS address ”Ref. 1”).
+ In addition, both pressure transducers have to be congured to cor-
responding VACUU•BUS addresses (VSK x and Ref. x). If the two ad-
dresses do not correspond (VSK x and Ref. y), no differential pressure
will be determined or displayed. Then it is necessary to recongure
the address of one pressure transducer to enable differential pressure
measurement (see section ”Reconguring the pressure transducer”
below).
➨ The displayed pressure value is: (”pressure value of Ref. x” minus
”pressure value of VSK x”) Torr
Reconguring the pressure transducer VSP 3000 or VSK
3000
Conguration of the MPT 100: See instructions for use of MPT 100.
Reconguring the VACUU•BUS address of a pressure transducer:
1. Switch the vacuum gauge DCP 3000 off.
2. Preferably connect only the pressure transducer to be recongured to
the DCP 3000.
3. Keep the ”VENT” key pressed and switch the DCP 3000 on (”ON/
OFF” key). Release the ”VENT” key.
4. Select the menu item ”Vacuubus” by turning and pressing the selec-
tion knob.
5. The rst component detected by the VACUU•BUS is displayed (e.g.,
VSK 1). If necessary, select the pressure transducer to be recong-
ured by turning and pressing the selection knob.
6. Press the ”VENT” key and keep pressed, then press additionally the
selection knob. A frame will replace the inverse display of the compo-
nent. Release the ”VENT” key.
7. Turn the selection knob until the address, to which you want to recon-
gure the pressure transducer (e.g., VSK 2 or Ref.1), is displayed.
8. Press the selection knob. If the conguration was successful, the menu
is left and the DCP 3000 will restart.
Comments to this Manuals