Type
PC 3002
VARIO
PC 3003
VARIO
PC 3004
VARIO
Maximum pumping speed (ISO
21360)*
m
3
/h 2.8 2.8 4.6
Ultimate vacuum (absolute) without
gas ballast**
mbar 7 0.6 1.5
Ultimate vacuum (absolute) with gas
ballast**
mbar 12 2 3
Maximum permissible inlet pressure
(absolute)
bar 1.1
Maximum permissible outlet pressure
(absolute)
bar
1.1
Maximum permissible pressure (abso-
lute) at gas ballast valve
bar 1.2
Permissible ambient temperature
storage / operation
°C -10 to +60 / +10 to +40
Permissible relative atmospheric
moisture during operation (no conden-
sation)
% 30 to 85
Power (electric) W 575
No-load speed min
-1
30 - 2400
Maximum permissible range of supply
voltage ( ±10% )
Attention: Observe specications of
rating plate!
100-120 V~ 50-60 Hz
200-230 V~ 50-60 Hz
Maximum nominal current draw at:
100-120 V~ 50/60 Hz
200-230 V~ 50/60 Hz
A
A
3.2
1.4
6.3
2.5
Device fuse slow blow fuse 4A (200-230V) / 8A (100-120V)
Motor protection thermal cutout
Degree of protection IEC 529 IP 20
Inlet hose nozzle NW 10 mm
Outlet hose nozzle NW 10 mm
Coolant connections (waste vapour condenser) hose nozzle NW 6-8 mm
Maximum permissible pressure of
coolant at waste vapour condenser
bar 6 (absolute)
Permissible range of coolant tempera-
ture
°C -15 to +20
Dimensions L x W x H approx. mm 422 x 258 x 516
Weight approx. kg 18.4 21.0
* Pumping speed of diaphragm pump
** Ultimate vacuum at setting ”Pump down” at speed ”HI”
We reserve the right for technical modications without prior notice!