Vacuubrand PC 3012 VARIO User Manual Page 40

  • Download
  • Add to my manuals
  • Print
  • Page
    / 75
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 39
Documents are only to be used and distributed completely and unchanged. It is strictly the users’ responsibility to check carefully
the validity of this document with respect to his product. Manual-no.: 999164 / 19/05/2009
page 40 of 75
Conguration
Adjustment 1013 mbar
RS-232...
Sensors...
Display...
Autostart Off
Defaults Cancel
- - - - - - - - Back - - - - - - - -
In the menu ”Conguration” the device parameters are preselected.
Adjustment: Adjustment of the pressure transducer under vacuum and/or at atmospheric pressure,
see also section ”Readjustment of CVC 3000”.
Adjustment at atmospheric pressure is carried out at an absolute pressure value between 1060 - 700
mbar and under vacuum at an absolute vacuum value between 0 - 20 mbar. In the range between 20
and 700 mbar no adjustment is possible; ---- mbar is displayed.
RS-232: Conguration of the interface, setting of parameters and commands, see section ”Inter-
face”.
Baud rate can be set to 19200, 9600, 4800 or 2400, parity on ”8-N-1”, ”7-O-1” or ”7-E-1”, Handshake
on ”no”, ”Xon-Xoff” or ”RTS-CTS” and remote on ”On” or ”Off”.
Sensors: Selection of the pressure transducer to be controlled (maximum 4 external sensors).
Display: Selection of the device parameters ”Brightness” between 0 - 100%, ”Contrast” between
0 - 100%, ”Sound” ”On” or ”Off”, ”Units” ”mbar”, ”hPa” or ”Torr”, ”Language” ”German”, ”English”,
”French”, ”Italian”, ”Spanish”, ”Turkish”, ”Korean”, ”Chinese”, ”Portuguese”, ”Russian”, ”Polish”,
”Dutch”, ”Japanese”
Autostart: If ”Autostart” is set to ”On” the controller restarts a running process automatically after a
mains failure. If this is unwanted, set Autostart to ”Off”.
Attention: If Autostart” is preselected, the process starts immediately after power failure without
pressing any further key. It is the users responsibility to ensure that no dangerous status of the sys-
tem due to the automatic start-up can occur and to provide appropriate safety measures. If neces-
sary, the user has to check prior to starting the process if the option ”Autostart” is enabled.
Defaults: If ”Defaults” is set to ”Load”, the controller is reset to factory set values. All stored programs
and parameters are deleted.
The screen-shot shows the factory-set values.
Preselections
Use the selection knob to select the parameters.
Function Conguration
Page view 39
1 2 ... 35 36 37 38 39 40 41 42 43 44 45 ... 74 75

Comments to this Manuals

No comments