Command Operation Response Description
IN_PV_1 current pressure
XXXX mbar/Torr/
hPa*
X.XXE±X mbar/
Torr/hPa*
unit according to preselections
IN_PV_Sx
current pressure
of pressure trans-
ducer x
XXXX mbar/hPa/
Torr*
X.XXE±X mbar/
Torr/hPa*
pressure of pressure transducer x
(order of numbering according to dis-
play in ”Sensors” menu)
IN_PV_X
current pressure
values of all con-
nected pressure
transducers
XXXX.X XXXX.X
...mbar/Torr/hPa*
X.XXE±X X.XXE±X
...mbar/Torr/hPa*
pressure values of all connected pres-
sure transducers
IN_PV_D data readout
XXX.XX:XX:XX X
...mbar/Torr/hPa*
read out data from DCP 3000 memory
IN_CFG
device set
preselections
yXXXXXX
X0XXXXX
X1XXXXX
X2XXXXX
XX0XXXX
XX1XXXX
XXX0XXX
XXX1XXX
XXXX0XX
XXXX1XX
XXXXXyX
XXXXXXy
y: 0......D: language
**
(hexadecimal)
pressure unit mbar
pressure unit Torr
pressure unit hPa
acoustic signal off
acoustic signal on
venting valve not connected
venting valve connected
fault indicator not connected
fault indicator connected
y: 1......8: number*** of active pressure
transducer
y: 1......8: pressure transducer quantity
IN_ERR fault status
0XXXX
1XXXX
X0XXX
X1XXX
XX0XX
XX1XX
XXX0X
XXX1X
XXXX0
XXXX1
no fault at venting valve
fault at venting valve
no overpressure
overpressure
no fault at pressure transducer
fault at pressure transducer
no external fault
external fault
last interface command correct
last interface command incorrect
IN_VER version DCP 3000 Vx.xx software version
IN_SP_1
time interval for
sending
XX:XX
mm:ss (minutes:seconds);
time interval for sending
IN_SP_2
time interval for
recording
XX:XX
mm:ss: (minutes:seconds);
time interval for recording