Readjustment
The vacuum gauge was adjusted using factory standards, which are traceable through regular calibra-
tion in an accredited laboratory (DAkkS calibration laboratory) to the national standard. Depending on
the process and/or accuracy requirements, check the adjustment from time to time and readjust if nec-
essary. For readjustment, the device has to be adjusted both at atmospheric pressure as well as under
vacuum.
Adjustment at atmospheric pressure
Ventilate the controller and/or the vacuum system. Make sure
that the vacuum connection at the controller is at atmospheric
pressure.
Select program ”Cal.Sens” at controller.
Use keys s and t to adjust the display to the actual local
atmospheric pressure.
Conrm value with key ”Enter”.
Note: To determine the actual atmospheric pressure, use an accurate barometer or get accurate read-
ing from the weather service, the next airport.......(take into account the difference in altitude between e.
g. airport and laboratory).
Adjustment under vacuum
Evacuate the controller via the vacuum connection to a pres-
sure < 0.5 mbar (e. g. by applying a good rotary vane pump).
Select program ”Cal.Sens” at controller.
The display is set to zero automatically.
Conrm value with key ”Enter”.
Note: Adjustment under vacuum with an actual pressure higher than 0.5 mbar reduces the accuracy of
the measurement. If the pressure is signicantly higher than 0.5 mbar, adjustment to a reference pres-
sure is recommended.
Adjustment at a reference pressure
Instead of adjustment under vacuum to a pressure < 0.5 mbar,
adjustment to a reference pressure within the range of 0 .....
20 mbar is possible.
Evacuate the controller via the vacuum connection to a refer-
ence pressure within the range 0 .... 20 mbar.
Select program ”Cal.Sens” at controller.
The display is set to zero automatically.
Use keys s and t to adjust the display to the actual refer-
ence pressure at the vacuum connection within the range
of 0 .... 20 mbar.
Conrm value with key ”Enter”.
Note: The accuracy of the value of the reference pressure will directly affect the accuracy of the adjust-
ment. If the nominal ultimate vacuum of a diaphragm pump is used as reference vacuum, the accuracy
of the controller might be doubtful. The diaphragm pump may not achieve the specied value (due to
condensate, poor state, failure of valves or the diaphragm).
1013mbar
0mbar
0....20mbar