Troubleshooting
Fault Possible cause Remedy
❏
No display.
➨
Mains cable not plugged in or
device not switched on?
✔
Plug in mains cable. Switch
device on. Check fuse in
building.
➨
Fuse defective?
✔
Check fuse in building and
fuse at switch output of VNC
2, replace if necessary.
➨
Other causes (device defec-
tive)?
✔
Contact local distributor.
❏
Display disappears, inter-
nal overload protection
activated.
➨
Thermal overload at control-
ler, ambient temperature too
high?
✔
Ensure adequate ventilation.
➨
Short circuit at connected
valves?
✔
Replace valves.
➨
Thermal overload at control-
ler, overload?
✔
Check current draw of con-
nected devices (pumps,
valves).
➨
Other causes (device defec-
tive)?
✔
Contact local distributor.
❏
Pressure reading incor-
rect.
➨
Device not adjusted correctly?
✔
Readjust controller.
➨
Humidity in pressure trans-
ducer?
✔
Let pressure transducer
dry, e.g. by pumping down.
Readjust if necessary. Detect
and eliminate cause.
➨
Pressure transducer soiled?
✔
See „Cleaning the pressure
transducer“, pg. 36.
❏
Controller does not
respond when pressing
keys.
➨
Controller set to remote
mode, ”P“ is displayed?
✔
Control controller via inter-
face or switch off remote
mode.
❏
Error message
”ErrValve”, two beeps.
➨
External venting valve re-
moved or defective?
✔
Connect or check venting
valve. In case install a new
venting valve or congure
controller without venting
valve. Switch controller off/
on.
❏
Error message
”ErrValve”, three
beeps.
➨
In-line valve removed or
defective?
✔
Connect or check in-line
valve. In case install a new
in-line valve or congure con-
troller without in-line valve.
Switch controller off/on.