Setting of the interface
Setting of the interface parameters directly at the controller is described below. Enter the device menu
Select, RS 232”. The factory-set values are underlined. Adjust interface parameters with keys s or
t and conrm with ”Enter”.
Baud 2400, 4800.9600 or 19200
Parity: 7 data bits odd (Parity 7-O-1); 7 data bits even (Parity 7-E-1);
8 data bits none (Parity 8-N-1)
Handshake (Handsh.): no Handshake (Handsh. None), XON/XOFF Handshake (Handsh. Xon-
Xoff), RTS/CTS Handshake (Handsh. RTS-CTS)
Remote On, Remote Off,
Startbit = 1, Stopbit = 1
Sending: timeout 1s, receiving: timeout 10s
In remote mode (Remote On, ”P” is displayed) all keys at the controller are without function. To return to
the manual operation of the controller, set the controller to the mode ”Remote Off” via the interface or
switch off controller and enter the device menu within the rst 2s after having switched on the controller
again.
Interface parameters
The controller VNC 2 is equipped with a serial interface (RS 232C, nine-pole Sub-D-plug).
Plug-in or remove the cable (cable RS 232C) from the interface only if the equipment is switched
off.
The interface is not electrically isolated from the measuring circuit.
For optimal electromagnetic compatibility assemble an interface lter (cat. no.: 638235).
The controller can be operated via serial interface. Measuring results, preselections and the status of the
controller can be read at any time.
Factory-set the read and write commands are completely compatible to the VACUUBRAND con-
troller CVC 2000. An extended instruction set compatible with CVC 3000 is available using the
command ”CVC 3”.
Calibration in the factory
The VACUUBRAND DAkkS calibration laboratory is accredited by the Deutsche Akkreditierungsstelle
GmbH (national accreditation body of the Federal Republic of Germany) for the measurable variable
pressure in the pressure range from 7.5*10
-4
Torr to 975 Torr (10
-3
mbar to 1300 mbar) in accordance with
the general criteria for the operation of testing laboratories dened in the DIN EN ISO/IEC 17025:2000
series of standards (accreditation number D-K-15154-01). The DAkkS is signatory to the multilateral
agreements of the European cooperation for Accreditation (EA) and of the International Laboratory Ac-
creditation Cooperation (ILAC) for the mutual recognition of calibration certicates.
Rely on calibration in the VACUUBRAND calibration laboratory:
- To meet the requirements of the DIN ISO 9000ff and 10012 series of standards regarding the calibra-
tion of inspection, measuring and test equipment at specied intervals.
- To document that the vacuum gauges calibrated are traceable to national standards of the PTB
(Physikalisch-Technische Bundesanstalt; German national institute for science and technology and
the highest technical authority of the Federal Republic of Germany for the eld of metrology and
certain sectors of safety engineering).
DAkkS Calibration Controller VNC 2 ...........................................................................................900217
Control of measuring equipment