Read commands compatible to ”CVC 3000”
2: RxD 7: RTS
3: TxD 8: CTS
4: DTR 9: +5V (Bluetooth)
5: Mass
Connector assignment
Description
unit according to preselections
maximum speed
(100% = ”HI”)
start-up pressure for VACUULAN or
two point control
delay
(00:00 = Off)
switch off pressure (”Maximum” for ”Vacuum con-
trol”, ”Minimum” for ”Continuous pumping”)
process runtime
00:00 = off
Response
XXXX mbar or
XXXX Torr or
XXXX hPa
XXX%
XXXX mbar or
XXXX Torr or
XXXX hPa
XX:XX hh:mm
XXXX mbar or
XXXX Torr or
XXXX hPa
XX:XX hh:mm
Function Command
IN_SP_1
IN_SP_2
IN_SP_3
IN_SP_4
IN_SP_5
IN_SP_6
set vacuum
speed
start-up
pressure
delay
switch off
pressure
A maximum of ten commands per second is possible.
Read commands and commands ”REMOTE”, ”CVC” and ”STORE” can be sent always. The sending
of other write commands is only possible, if ”Remote On” is selected.
The commands have to be written in capital letters.
Command and parameter have to be separated by a blank.
The string is terminated with <CR> or <CR><LF>.
The answer of the controller is always terminated with <CR><LF>.
Numerical values and parameters can be written without leading zeros.
The answer of the controller is always with leading zeros.