Abbreviations
FPM: Fluoroelastomer
GND: Ground
PE: Polyethylene
PTFE: Polytetrauoroethylene
PWM: Pulse width modulation
Position Component
1 Power connection
2 ON/OFF switch
3 Voltage selection switch
4 Fuse holder
5 Pump rating plate
6 Inlet
7 Outlet
8 Fan
9 Handle (removable) / recessed grip
10 Control line
Pump parts
Gas inlet temperatures
* if pumping potentially explosive atmospheres: 50 °F to 104 °F (+10°C to +40°C)
Operating condition Inlet pressure
Permitted range of gas
temperatures at inlet
Continuous operation
> 75 Torr (100 mbar)
(high gas load)
➨ 50 °F to 104 °F
(+10°C to +40°C)
Continuous operation
< 75 Torr (100 mbar)
(low gas load)
➨ 32 °F to 140 °F*
(0°C to +60°C*)
Short-time
(< 5 minutes)
< 75 Torr (100 mbar)
(low gas load)
➨ 14 °F to 176 °F*
(-10°C to +80°C*)