Fault Possible cause Remedy
Pump too noisy.
Atmospheric or high
pressure at the pump
inlet?
Connect hose or silencer
to pump outlet. Be careful
not to cause outlet over-
pressure, especially with
condensable vapors.
Diaphragm crack or
diaphragm clamping
disc loose?
Perform maintenance.
Other than above men-
tioned causes?
Contact local distributor.
Pump seized.
Contact local distributor.
A service manual with exploded view drawings, spare parts list and
directions for repair is available on request.
+ The service manual is intended for trained service people only.