Fault Possible cause Remedy
Pump does not
start or stops im-
mediately.
Electrical power cord
not plugged in, electri-
cal supply failure?
Plug in power cord.
Check fuse. Check elec-
trical supply voltage.
Device fuse blown (if
applicable)?
Identify cause of failure.
Replace device fuse.
Overpressure in outlet
line?
Remove blockage in line,
open valve, or reduce
overpressure.
Motor overloaded?
Allow motor to cool down,
identify and eliminate
cause of failure. Manual
reset is necessary. Switch
off pump or unplug.
Pump does not
achieve its ultimate
vacuum or usual
pumping speed.
Centring ring at small
ange connection not
correctly positioned, or
leak in the pipeline or
vacuum system?
Check pump directly -
connect vacuum gauge
directly at pump inlet -
then check connection,
pipeline and vacuum
system if necessary.
Long, narrow vacuum
line?
Use lines with larger di-
ameter, length as short as
possible.
Pump has been ex-
posed to condensate?
Allow pump to run for
some minutes with atmo-
spheric pressure at the
inlet to purge.
Deposits have been
formed inside the
pump?
Clean and inspect the
pump heads.
Diaphragms or valves
damaged?
Replace diaphragms and/
or valves.
Outgassing substances
or vapor generated in
the process?
Check process parame-
ters.
Only VARIO-SP ver-
sion: Pump tempera-
ture too high (motor
speed reduced)?
Ensure sufcient cooling
of the pump or reduce
inlet pressure.
Troubleshooting