When assembling, ensure vacuum-tightness. After as-
sembly, check the whole system for leaks.
Use a suitable valve to isolate the pump from the vacuum
application. This is to allow the pump to warm up before
pumping condensable vapors and to clean the pump after
use before it is switched off.
NOTICE
Connecting the outlet
Outlet via silencer.
Attention: Use the silencer only in case of low
gas throughput. Check the silencer regularly
for permeability! In case, install a hose nozzle
instead (see „Accessories“, pg. 44).„Accessories“, pg. 44).).
Connecting an exhaust line at the pump outlet:
Unscrew the silencer and install a hose noz-
zle DN 6 mm or a hose nozzle DN 10/6 mm
(thread G1/8”).
If dangerous or polluting uids could be released at
the outlet, install an appropriate system to catch and
dispose of those uids.
NOTICE
outlet
+ Connect a gas-tight exhaust line at the pump
outlet (see above) if necessary. Always vent
exhaust gases appropriately (e.g., into a
fume hood).
+ Never block the gas outlet. The exhaust
line must always be free of obstructions (no
back pressure) to ensure an unimpeded
discharge of gas. The cross-section of the
outlet tubing must be at least the size of the
pump’s exhaust connection.
Reduce the transmission of vibration. Prevent mechan-
ical load due to rigid pipelines. Insert elastic hoses or
exible elements as couplings between the pump and
rigid pipes.